IM-105172 notice for IM 2020 Edition (v8.17.XX) Instrument Manager (IM) customers

Conseils et astuces

Data Innovations (DI) is raising awareness to Known Issue IM-105172. This issue impacts customers on IM 2020 and subsequent minor releases for this edition.

When an internal message is created and routed to a connection that has been previously deleted, QMGR goes into an Error state and files the following error: imrouter+9 ^imrouter^imconfig(“line”,#) (Job# #)

Before deleting a connection, please review the details of Known Issue IM-105172 in My DI Community or in Customer Web Portal (CWP) and perform these preventive steps:

  • If using Specimen Routing, remove the connection from the Destination column.
  • If using the connection in a driver configuration, remove that connection from the driver configuration.
  • If using a cluster, remove the connection from the cluster.
  • If you have a cluster that has no connections associated to it, be sure to either remove that cluster or add connections to it.
  • If the connection you wish to delete is enabled as a destination line for a separate connection, the connection must be removed on that connection’s Destination Lines.
  • Check rules that may route an order or rerun to that connection. If rules are using that connection, remove that referenced connection from the rule.
  • Remove the connection from groups that have access (under Security > Users and Groups).
  • Instead of deleting the connection, rename the connection to prevent the LIS from sending orders to it.

This Known Issue was corrected in v9.00.00.

If the error has already happened, please contact your Business Partner (if applicable) or the appropriate DI Regional Support team to help resolve the issue.

More Tips & Tricks Articles

Conseils et astuces

TCP/IP Keep-Alive

Data Innovations’ (DI’s) Instrument Manager (IM) offers connectivity that enables integration with virtually any instrument, LIS, or EMR/EHR regardless of...

Instrument Manager, Tips & Tricks

Following Driver Updates in My DI Community

Aperçu

Data Innovations (DI) provides ways to stay up to date on the latest driver updates.The first way is right through...
Instrument Manager, Tips & Tricks

Net Gateway Startup Issues (Error 5023)

Aperçu

When IRIS and Instrument Manager (IM) are starting after a server shuts down or restarts, the .NET gateway is not...
Conseils et astuces

How to Install Instrument Manager™ Core Software

NOTE These instructions apply for Data Innovations’ North American customers. If this is your first time installing IM, please contact Data...

Conseils et astuces

How to help Data Innovations help you when requesting Support

Data Innovations' Tips & Tricks are intended to promote the effective and productive use of Instrument Manager. Fortunately, Instrument Manager...

Conseils et astuces

Evaluation Rules for Clinical Ranges Precaution

When building an Autoverification Rule Set it's common to include evaluation rules for Clinical Ranges. These rules typically set test...

Conseils et astuces

Instrument Manager System Preventative Maintenance (volume 1)

Instrument Manager is a robust software application with a proven track record of operational stability. However, as with any software...

Conseils et astuces

Instrument Manager System Preventative Maintenance (volume 2)

This month's Tips & Tricks is the second installation of Preventative Maintenance recommendations intended to help you maintain a stable...

Vous souhaitez en apprendre davantage ?